Operation Precautions
Before Connecting a Logic Probe to the Measurement Object
DANGER
To avoid electric shock and short circuit accidents or damage to the instrument,
confirm the following:
•
The ground pin in the LOGIC connector (plug) of Models 9320-01 and 9327
Logic Probes is not isolated from the instrument’s ground (common ground).
Use grounding-type polarized power cords for the measurement object and this
instrument, and obtain power from the same mains circuit.
Connecting to different mains circuits or using a non-grounding power cord
may cause damage to the measurement object or this instrument because
of current flow through the logic probes resulting from potential difference
between the grounds of the different wiring systems.
To avoid these problems, we recommend the following connection procedure:
Measurement
Object
Measurement
Object
GND
Logic probe
Logic probe
Functional Earth
Terminals
Memory
HiCorder
Memory
HiCorder
Connect this instrument to the
same outlet as the measurement
object using the (supplied)
grounding polarized power cord.
Connect the measurement object’s
ground to the GND terminal
(functional earth terminal) of this
instrument.
(Always obtain power from the
same mains circuit.)
See “2.5.2 Ground the Functional
Earth of This Instrument” (page
Before Turning the Power Supply On
WARNING
•
To avoid electrical accidents and to maintain the safety specifications of this
instrument, only connect the provided power cord to an outlet.
•
Before turning the instrument on, make sure the supply voltage matches that
indicated on its power connector. Connection to an improper supply voltage
may damage the instrument and present an electrical hazard.
CAUTION
Avoid using an uninterruptible power supply (UPS), DC/AC inverter with rectangular
wave or pseudo-sine-wave output to power the instrument. Doing so may damage the
instrument.
To avoid damaging the power cord, grasp the plug, not the cord, when unplugging it
from the power outlet.
Turn off the power before connecting or disconnecting the power cord.
10
9
8
7
6
5
4
3
2
1
Appx.
Index
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......