190
Analog Signal Triggers
9.3 Analog Signal Triggers
The process below explains how to set up the different types of
analog triggers. These settings can be
configured on the Trigger settings window (
[Analog Trg.]
sheet).
Procedure
Press the
DISP
key to open the Waveform screen, and then press the
TRIG.SET
key to open the trigger settings
window (
[Analog Trg.]
sheet).
1
3
1. Level Trigger (p. 191)
2. In-Window Trigger
Out-of-Window Trigger (p. 191)
3. Voltage Sag Trigger (p. 192)
4. In-Period Trigger,
Out-of-Period Trigger (p. 192)
5. Glitch Trigger (p. 193)
1
Move the cursor to
[Type]
for the channel for which you want to change the
settings.
2
Select the trigger type with the
F
keys.
3
Move the cursor to the parameter item with the
CURSOR
keys.
4
Select the parameter value with the
F
keys.
To copy the setting to another channel
The Trigger settings window ([Analog Trig] sheet) can be used to copy a setting.
Refer to “8.9 Copy Settings to Other Channels (Copy Function)” (p. 163).
Analog triggers cannot be set when the FFT function is used and
[Reference]
is set to
[From Memory]
on the [Status] sheet.
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......