354
Other Functions
Sequential saving
Possible by specifying start and end blocks
Sequential saving
dead time
When block display is OFF: 1 to 8 samples (timebase: 5 μs/div to 20 μs/div)
1 sample (timebase: 50 μs/div or more)
When block display is ON: 40 ms or more (timebase 5 μs/div to 20 μs/div)
1 sample (timebase: 50 μs/div or more)
Multi-block saving
Saving of waveforms is possible by specifying the desired blocks of divisions
Block display
Display can be set to ON or OFF
Block overlaying
Desired blocks or all blocks
Cursor Measurement Functions
Cursor functions
All functions supported
Number of cursors
Two (Cursors A and B)
Cursor types
Line (vertical or horizontal), trace
Cursor movements
A Cursor, B Cursor, A&B Cursors
Measurement
functions
A Cursor: Potential at each cursor, time from trigger event
A&B Cursors: Time differences between cursors, potential, frequency (period)
Cursor-supported
channels
All channels (default), any specified channel
Included functions
Partial printout*, partial saving
* When Model U8350 Printer Unit is installed
Monitor Functions
Monitor display
Level monitor: Display the level and numerical value by selecting the level monitor with
the
DISP
key
DMM display: Display the numerical values by pressing the
DISP
key
Numerical value
display
Instantaneous value display, with hold function
Level monitor
sampling
10 kS/s (fixed)
Refresh rate
0.5 s or more, depends on the NPLC settings when using the Model MR8990 Digital
Voltmeter Unit
Position Display (VIEW) Functions
Display function
All types of positions
Memory division blocks (When using memory division)
Waveform search results
Past waveform history (When memory division is OFF, depends on recording length)
Position display
Displays the relative position of the currently displayed waveform within the overall
recording length, as well as locations of cursors and trigger events
Block display
Block usage status (When memory division is ON)
Past waveform history status (When memory division is OFF)
Displayed block position
Jump function
Jump to trigger position/cursor position
Jump to desired block (When using memory division)
Jump to past waveform (When memory division is OFF)
Jump to waveform search position
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......