343
External I/O
17.2.4 Trigger Output (TRIG OUT)
You can output a signal when a trigger event occurs. This feature allows multiple instruments to be
controlled for parallel synchronous operation.
Signal Output Procedure
1
Connect the cables for the output signals to the TRIG OUT and GND terminals.
Refer to “17.1 Connect to External Control Terminals” (p. 337).
2
Press the
SYSTEM
key to open the
[Environment]
sheet, and move the cursor to
[TRIG
OUT]
.
3
Select the output signal type for the trigger output terminal.
Pulse
After LOW level output, the signal goes HIGH after a specified interval.
Level
After the trigger occurs, a LOW level signal will be output during waveform
retrieval.
When a trigger event occurs, a pulse wave changing from HIGH (4.0 V to 5.0 V) to LOW (0 V to 0.5
V) will be output from the TRIG OUT terminal.
Output signal
Open drain output (with voltage output) active LOW*
Output voltage range
HIGH level: 4.0 V to 5.0 V
LOW level: 0 V to 0.5 V (current: 15 mA)
Pulse width
Pulse width at pulse setting time: 2 ms ± 1 ms
Pulse width at level setting time: (sampling rate × no. of data points after
trigger) or more
Maximum input voltage
50 V DC 50 mA 200 mW
* Triggering should occur when the signal voltage level changes from HIGH to LOW.
HIGH
4.0 V to 5.0 V
LOW
0 V to 0.5 V
GND
TRIG OUT
100
kΩ
10
kΩ
5 V
1 ms or
greater
• Trigger events occur and signals are output when the auto-ranging function is used with the Memory
Function. Keep this in mind when performing auto-range measurement while using the trigger output
terminal.
• When using memory division, the trigger output (TRIG_OUT terminal output) may output the Low
level or output erratically under the following conditions.
•
The timebase is 5 μs/div to 100 μs/div
• The record (measurement) time is 5 ms or less
• Tracking wave display is
[Off]
.
17
External Control
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......