62
Select Input Channels
• When input coupling is set to GND, the waveform will have no amplitude and you cannot set the
range.
•
Due to the influence of filter attenuation, correct range setting may not be possible.
•
When choosing trigger settings, set the vertical axis (voltage axis) range first. If the range is changed
after specifying the trigger, the trigger setting may change.
•
When using the Variable function, set the vertical axis (voltage axis) range first. If you change the
range setting after applying the variable function observation with sufficient precision may not be
possible.
•
When using the Variable and Scaling functions together, apply the Scaling function first. If you apply
the Scaling function after the Variable function, you may not get the intended display result.
The setting workflow for logic channels (standard LOGIC terminals LA - LH, expansion LOGIC terminals
L1A to L8D) is explained below.
1 Select screen display related settings
Refer to...
Set logic recording width
2
Waveform display color and position settings
Refer to...
Set waveform display position
Set waveform display color
•
Waveform display position can be specified in 1% increments.
• Not displayed for X-Y1 and X-Y4 screens.
• Up to two logic units can be installed as units 1 to 8. Logic units installed in other positions cannot be
used.
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......