246
FFT Analysis Condition Settings
13.3.2 Select the
Data Source for Analysis (Reference Data)
Select the data to be used for FFT analysis.
There are two analysis methods: analysis using new measurements and analysis of data measured using
the memory function.
Procedure
Press the
STATUS
key to open the Status screen, and then select the
[Status]
sheet.
To set from the Waveform screen, refer to (p. 261).
1
Select the reference data.
Move the cursor to
[Reference]
.
New Data
Acquire a new waveform for analysis.
From Memory Calculates data measured using the
memory function.
1
2
When the input data
[Reference]
is
[From Memory]
Analysis is performed until the
specified number of FFT analysis
points have been processed, then
the data is shifted by that amount
and analysis repeats until all of the
previously acquired data has been
processed. (If the amount of data is
less than the specified number of FFT
analysis points, no analysis occurs.)
When no trace is displayed after
pressing the
START
key
Analysis is impossible if
[From
Memory]
is selected as the input data
source and no recorded data exists in
the instrument's memory.
Either select
[New Data]
as the input
data source, or load the data to be
analyzed before pressing the
START
key again.
2
When you have finished choosing settings, press the
START
key
When using
[New Data]
Measurement starts to acquire data for the number of
analysis points specified as the
[Sampling Point]
, and
FFT analysis is performed.
When using
[From Memory]
Analysis is performed on the number of specified points
from data previously recorded in memory (Memory function
data).
The analysis starting point can also be specified.
Refer to “13.8.1 Analyze After Specifying an Analysis
The frequency range is selected automatically.
Refer to “Relationship Between Frequency Range,
Resolution and Number of Analysis Points” (p. 248).
When
[Reference
] is set to
[From Memory]
, the frequency is automatically set.
The setting cannot be changed.
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......