333
Controlling the Instrument with Command Communications (LAN/USB)
16.7
Controlling the Instrument with Command
Communications (LAN/USB)
You can control the instrument remotely via the communications interface (LAN or USB).
• For details, see the Communication Command Instruction Manual on the supplied application disc.
• Before using command communications, the LAN settings or USB settings and connection must have
been configured properly.
About LAN, refer to ““16.1” (p. 310); USB“16.6” (p. 328).
CAUTION
In an environment where noise from inverters or similar devices is present, errors
may occur when controlling the instrument remotely. Take care to ensure that the
environment is not subject to excessive noise.
Setting Details
Delimiter
The Delimiter item specifies LF or CR+LF as the newline delimiter in command response
messages. The instrument understands both settings: LF or CR+LF.
Header
Use for control of communications commands.
The Header item specifies whether to prefix headers to command response messages.
For more information about commands, refer to the Communications operation manual on the
supplied CD.
Command Port
(Port number) (LAN
only)
The instrument uses the TCP/IP protocol for communications. TCP/IP allows communicating
devices to establish multiple connections, which are distinguished by port number. By default
the instrument uses port numbers 8800 to 8809.
•8800 to 8801 reserved
•8802 (instrument is server): For communications command control
•
8803 to 8809 reserved
Normally these ports do not need to be changed. You can change them if certain ports cannot
be used for security reasons, or if certain ports are not available on the communicating PC.
Set only the most significant three digits. The least significant digit (0 to 9) is used by the
instrument, or reserved for use by the instrument.
16.7.1
Settings on the Instrument
Set items related to command communications.
Procedure
Press the
SYSTEM
key to open the System screen, and then select the
[Interface]
sheet.
1
Set the delimiter.
Move the cursor to
[Delimiter].
LF
Send character code 0x0a.
CR+LF
Send character codes 0x0d and 0x0a.
2
Select header settings.
Move the cursor to
[Header].
Off
Do not add a header to response data.
On
Add a header to response data.
3
Set the communications command port. (LAN only)
Move the cursor to
[Port Number]
and enter the port number.
16
Connect to a PC
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......