54
Measurement Configuration
3.3.2
Timebase and Sampling Rate
The timebase setting establishes the rate of input signal waveform acquisition, specified as time-per-
division on the horizontal axis (time/div).
The sampling setting specifies the interval from one sample to the next.
(The sampling rate is shown in brackets under the timebase for the Memory function
(see illustration at right). The displayed sampling rate changes as the time axis range
changes.)
Procedure
Press the
DISP
key to open the Waveform screen.
With the Memory Function
1
Move the cursor to the
[Timebase]
item.
2
Set the time per division on the horizontal axis
(time axis).
5 (
Default setting
), 10, 20, 50, 100, 200, 500 μs/
div
1, 2, 5, 10, 20, 50, 100, 200, 500 ms/div
1, 2, 5, 10, 30, 50 s/div, 1 min/div, 100 s/div
2, 5 min/div
To control sampling with an external signal, select
[External]
and input the signal from the
external sampling terminal. To reset the setting to the internal sampling, select
[Internal]
and set
the time.
When external sampling is selected, the number of samples per division can be set in the range
from 10 - 1000 S/div.
“17.2.3 External Sampling (EXT.SMPL)” (p. 342).
With the Recorder Function
1
Move the cursor to the
[Timebase]
item.
2
Set the time per division on the horizontal axis
(time axis).
10 (
Default setting), 20, 50, 100, 200, 500 ms/div
1, 2, 5, 10, 30, 50 s/div, 1 min/div, 100 s/div
2, 5, 10, 30 min/div, 1 h/div
3
Move the cursor to the
[Sampling]
item.
4
Set the sampling rate.
1 (
Default setting
), 10, 100 μs
1, 10, 100 ms
(Select a sampling rate less than 1/100 of the time
base)
The range of choices depends on the selected timebase.
The higher the sampling rate, the more detailed the changes that can be observed.
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......