57
Measurement Configuration
3.3.3
Recording Length (Number of Divisions)
Set the length (number of divisions) to record each time data is acquired.
Procedure
Press the
DISP
key to open the Waveform screen.
With the Memory Function
1
Move the cursor to
[Shot].
2
Select the type.
Fixed Shot
Select from a range of preset values.
User Shot
Freely specify any value in 1-division units.
3
Set the recording length.
(Fixed Shot)
25, 50, 100, 200, 500, 1,000, 2,000, 5,000, 10,000, 20,000,
50,000, 100,000 div
200,000 div (in 4/8/16 ch mode)
500,000 div (in 4/8 ch mode)
1,000,000 div (in 4 ch mode)
(User Shot)
1 to 160,000 div (in 32 ch mode)
1 to 320,000 div (in 16 ch mode)
1 to 640,000 div (in 8 ch mode)
1 to 1,280,000 div (in 4 ch mode)
With the Recorder Function
1
Move the cursor to
[Shot].
2
Select the type.
Fixed Shot
Select from a range of preset values.
User Shot
Freely specify any value in 1-division units.
3
Set the recording length.
(Fixed Shot)
Cont.
On-Off
When
[On]
is selected, measurement is
carried out while continuously overwriting the
data in memory.
Data from the point at which
measurement was stopped back to the maximum
recording length can be saved or printed.
25, 50, 100, 200, 500, 1,000, 2,000, 5,000, 10,000, 20,000,
50,000 div
(User Shot)
1 to 80,000 div
3
Measurement
Содержание MR8827
Страница 19: ...14 Operation Precautions ...
Страница 81: ...76 Start and Stop Measurement ...
Страница 111: ...106 Manage Files ...
Страница 125: ...120 Miscellaneous Printing Functions ...
Страница 143: ...138 View Block Waveforms ...
Страница 191: ...186 Setting Output Waveform Parameters ...
Страница 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Страница 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Страница 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Страница 295: ...290 FFT Analysis Modes ...
Страница 309: ...304 Editor Command Details ...
Страница 387: ...382 Module Specifications ...
Страница 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Страница 431: ...A26 FFT Definitions ...
Страница 436: ......