07-19-01-518 – January 2009
4B-19
Cassette Properly Installed in Cassette Compartment
25. Place the donor and anticoagulant lines securely in the
channels directly in front of the donor safety clamp. The air
detectors are located within the left channel.
Donor and AC Lines Placed in Air Detectors
WARNING: The air detector housing and tubing channel
must be dry for proper function.
26. Check all tubings to ensure they are clear of the cassette
door latches.
27. Close the cassette compartment door and ensure it is secure.
WARNING: Do not operate the instrument if the cassette
compartment door is open. Do not operate the instrument if
the cassette compartment or pump block gasket is damaged.
Cassette
Guides
Cassette
Guides