A
LYX
C
OMPONENT
C
OLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
C
HAPTER
4C
–
A
LYX
2RBC-NLR
C
OLLECTION
P
ROCEDURE
4C-28
07-19-01-518 – January 2009
After all donor information has been entered, the
red cell volume
button
will display the default procedure volume as specified by
the procedure nomogram (located in the Appendix) or center
requirements.
14. To adjust the red cell volume, touch the
red cell volume button
.
Touch the button that displays the desired red cell collection
volume for the procedure.
Note:
Target volumes available for the A
LYX
2RBC-NLR
procedure are 360 mL and 400 mL.
Typical Select Red Cell Product Volume Overlay
15. Touch the
check button
to return to the Main Data Entry
Screen once the collection volume has been verified on the
screen overlay.
WARNING: Base the volume of the product(s) to be
collected on the maximum allowable target collection
volume(s) and on sound medical judgment, conforming to
applicable regulations and guidelines set by the appropriate
health regulating agencies.
Note:
Red cell target collection may have restricted volume
requirements depending on local guidelines.
Note:
The instrument will not allow entry of donor or product
inputs inconsistent with configured settings of the instrument’s
administrative settings. An audible tone and flashing data button
indicate entry of unacceptable data input.
Red Cell
Volume
Button
Check Button
Target
Volume
Button
Target
Volume
Button