A
LYX
C
OMPONENT
C
OLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
C
HAPTER
5
–
T
ROUBLESHOOTING
07-19-01-518 – January 2009
5-9
Typical Non-Recoverable Alarm with
Remove Kit Icon
A non-recoverable alarm that displays the
remove kit icon
only
indicates a condition where the procedure must be ended. The
operator should follow these steps to end the procedure:
1. Press the STOP button.
2. Touch the
end button
.
3. Follow the instructions and/or answer the questions that are
displayed on subsequent screens to proceed with ending the
procedure. In some instances, fluid may be returned to the
donor and/or the product may be processed.
Note:
Refer to Chapter 3 for additional instructions on using the
STOP button.
WARNING: If the instrument is accidentally turned OFF
during a procedure, immediately isolate the donor from the
instrument by closing the clamp on the donor line. Do not
turn the instrument ON with the donor connected.
Note:
Any products obtained under these circumstances should
not be used as validation or quality control samples.
Remove Kit Icon
End Button