A
LYX
C
OMPONENT
C
OLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
C
HAPTER
4A
–
A
LYX
2RBC-LR
C
OLLECTION
P
ROCEDURE
07-19-01-518 – January 2009
4A-53
6 - Whole Blood Processed -
the amount of donor whole blood
that was drawn and separated during the collection process,
expressed in mL.
Note:
The volumes defined for items 7, 8, and 9 appear in two
columns on the screen. The volumes in the left-hand column apply
to the storage container placed on the plasma scale. The volumes
in the right-hand column apply to the storage container placed on
the in-process scale.
7 -
Red Cell Product Volume -
the respective total volumes of
each red cell storage container, including red cells, anticoagulant,
plasma, and red cell preservation solution, expressed in mL.
8 -
Red Cell Preservation Solution Volume -
the respective
volume of red cell preservation solution found in each red cell
storage container, expressed in mL.
9 -
Anticoagulant Volume -
the respective volume of
anticoagulant found in each red cell storage container, expressed in
mL.
10 - Saline Used -
the total amount of saline removed from the
saline container for prime, return phases, and any extra saline given
to the donor during the procedure, expressed in mL.
11 - Anticoagulant Used -
the total amount of anticoagulant that
was removed from the anticoagulant container during the
procedure, expressed in mL.
12 - Filtration Time -
the total amount of time used to filter the
red cells and preservation solution, expressed in minutes:seconds.
13 - **Post-Leukoreduction Product Recovery
– the volume of
product recovered in the red cell storage containers after filtration,
expressed as a percentage.
14 - Procedure View Button –
Provides access to
summary
tabs
for the donor, procedure results, kit, and any alarms that occurred
during the procedure.
Note:
The product recovery displayed on this screen is based on
the total volumes of the products, not on absolute red cell
recovery.
** The Post-Leukoreduction Product Recovery percentage may or
may not be displayed according to the Software Version or settings
of your A
LYX
instrument. Refer to the Calculating Percent
Содержание 4R5707