![Fenwal 4R5707 Скачать руководство пользователя страница 385](http://html.mh-extra.com/html/fenwal/4r5707/4r5707_operators-manual_553093385.webp)
A
LYX COMPONENT COLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
C
HAPTER
5
–
T
ROUBLESHOOTING
07-19-01-518 – January 2009
5-35
Checking/Replacing Fuses
1. Ensure the instrument is NOT plugged into an outlet.
2. Disconnect the instrument’s power cord from the back of
the instrument.
3. With a flat-head screwdriver, gently remove each fuse (2)
located just above the power cord connection site. Refer to
the following figure.
Removing Fuses from the Instrument
4.
Inspect both fuses to determine if they appear functional. A
non-functional fuse may have a broken coil or burn marks
on the inside. The fuses are easily removed from their
holder.
Fuse Removed from Holder
5. Replace any fuse that looks like it may be non-functional
with the same type of fuse. Refer to “Electrical
Specifications” in Chapter 7 for the correct fuse.
Содержание 4R5707