A
LYX COMPONENT COLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
C
HAPTER
5
–
T
ROUBLESHOOTING
5-32
07-19-01-518 – January 2009
Code
Icons
Meaning
Resolution
11057
Air in Line
The system has detected air in the
donor line.
See Alarm 7025.
11058
-
10060
Non-Recoverable
Alarm Screen
The system has detected an issue
with the power supply voltage.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
11061
Non-Recoverable
Alarm Screen
The system has detected an issue
with the STOP button.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
11062
&
11063
Non-Recoverable
Alarm Screen
The system has detected an issue
with the donor pump.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
11064
Non-Recoverable
Alarm Screen
The system has detected an issue
with the cassette door. The door
may not be closed.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
11066
Non-Recoverable
Alarm Screen
The system has detected that the
instrument is in Test Mode.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
11067
Non-Recoverable
Alarm Screen
The system has detected air in the
donor line that cannot be cleared.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
11068
Non-Recoverable
Alarm Screen
The system has detected an issue
with the pressure cuff.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
11069
Non-Recoverable
Alarm Screen
The system has detected that the
instrument is in The Hotline Mode.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
11070
Non-Recoverable
Alarm Screen
The system has detected potentially
coagulated blood.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
11071
Non-Recoverable
Alarm Screen
The system has detected an issue
with the pneumatic circuit.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
11072
Non-Recoverable
Alarm Screen
The system has detected an issue
with the pneumatic valves.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
11075
Non-Recoverable
Alarm Screen
The system has detected an issue
with the pneumatic valves.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
12020
-
12025
Non-Recoverable
Alarm Screen
The system has detected a software
issue.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
13006
Scale has been
Disturbed
The system has detected weight on
a scale when it expects no weight
during the QC check.
- Check the scale that is highlighted on the
screen.
- Refer to Scale Calibration section in Chapter 3
for more information.
Содержание 4R5707
Страница 2: ......