Hanging Split Products to Weigh
6.
Hang one satellite bag on the plasma scale and the other on
the red cell scale. Open the clamps to distribute plasma
equally between the two containers, if desired close clamps.
7.
Seal the tubing on both satellite plasma containers
immediately next to the “Y”. Disconnect the “Y” and
remove the clamps.
8.
Allow the plasma product volume readings to stabilize.
9.
Touch the
check button
to confirm that the products have been
hung.
Typical Final Split Products to be Weighed Screen
10. A Task Confirmation Screen will be displayed to remind the
operator of the tasks that need to be completed prior to
finalizing the weights.
Task Confirmation Screen - Split Plasma Product Volumes
Check Button
Back Button