A
LYX
C
OMPONENT
C
OLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
C
HAPTER
4B
–
A
LYX
RBC/P
LASMA
C
OLLECTION
P
ROCEDURE
4B-14
07-19-01-518 – January 2009
Installing the Separation Chamber
Depending on the centrifuge carrier on your A
LYX
instrument,
one of two configurations will be available. The following
instructions for Steps 11 through 14 should be used if the carrier
on your instrument has a latch with a tab that must be opened to
insert the separation chamber and closed to lock it in the correct
position. This configuration is shown below.
11. Rotate the centrifuge carrier so that the two umbilicus guides
are on the left and the latch is facing up. Open the latch by
pressing on the tab and pulling up on the latch.
Centrifuge Assembly with Latch Closed
Latch in Open Position
12. With the umbilicus pointing toward the front of the
instrument, slide the separation chamber into the carrier.
Latch
Carrier
Groove
Latch
Tab
Umbilicus
Guides