A
LYX
C
OMPONENT
C
OLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
C
HAPTER
4B
–
A
LYX
RBC/P
LASMA
C
OLLECTION
P
ROCEDURE
4B-22
07-19-01-518 – January 2009
CAUTION:
If the disposable kit is not primed, it should be
used within 8 hours after being loaded on the instrument.
However, once the cannulae have been broken on a closed kit, or
once kit-checks have started on a functionally closed kit, a
procedure must begin within 1 hour. After this time, the
instrument will alarm and prompt the operator to remove the kit.
The operator must press the STOP button, end the procedure
and remove the kit. A new procedure may then be started. (See
Chapter 3, Automatic Cassette Refresh and Kit Timer for
additional information.)
CAUTION:
A disposable kit may only be reloaded as long as
the cannulae have not been broken, saline has not been spiked, if
applicable, and the sterility barrier(s) are intact. Otherwise if the
one-hour limit is exceeded, the kit must be discarded.
Note:
After a kit is loaded, minor adjustments may be made to
re-position the instrument as long as it remains plugged in with
the power ON.