PERATING CONSIDERATIONS
3-10
07-19-01-518 – January 2009
Note:
Procedure screens used in this manual are for general
illustration. These screens and actual instrument screens may
vary.
Typical Stop Test Screen
The instrument has been installed properly if the fans are
running and the Stop Test Screen is displayed after the A
LYX
Logo Screen is displayed. If the instrument has recently been
exposed to conditions outside of its normal operating
requirements, as outlined in Chapter 7 of this manual, allow the
instrument to remain running with the Stop Test Screen
displayed for at least one hour prior to continuing. This will
allow the instrument to equilibrate.
If the instrument fails to perform as expected to this point, call
a qualified service representative.
CAUTION:
The touch screen is the main input device for the
instrument. Do not touch the screen during operation except
to interact with the instrument.
CAUTION:
Do not operate the instrument unless both fan
filters are properly installed.