A
LYX
C
OMPONENT
C
OLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
C
HAPTER
5
–
T
ROUBLESHOOTING
5-12
07-19-01-518 – January 2009
Recoverable Alarms
Recoverable alarms may indicate conditions that affect donor
safety, donor comfort, or product quality if not resolved. These
alarms may occur during instrument set-up, kit installation, while
a donor is connected, or during the processing phase. Operator
attention is required.
Persistent recoverable alarms may require that the procedure be
ended. If you experience a persistent alarm situation, end the
procedure by pressing the STOP button, following the
instructions/choices on subsequent screens, complete the
end-procedure process, and contact the Hotline at your earliest
opportunity.
During a recoverable alarm, the instrument:
•
Automatically implements a safe-state mode in which the
donor is isolated. Depending on the alarm, separation may or
may not stop.
•
Displays a continuous yellow signal light at the top of the
instrument.
•
Sounds an audible signal.
•
Displays an alarm overlay on the touch screen.
When a recoverable alarm occurs, the instrument provides an
alarm screen that provides information to help resolve the
problem. Refer to the following figure.
Typical Recoverable Alarm Screen
Alarm
Code
Time
Alarm
Occurred
Date
Help
Button
Mute
Button
Alarm
Icons
Clear Line
Button
Cuff
Pressure
Button
Check
Button
Possible
Location of
Problem