A
LYX
C
OMPONENT
C
OLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
C
HAPTER
5
–
T
ROUBLESHOOTING
5-4
07-19-01-518 – January 2009
Non-Recoverable Alarms Requiring Power Cycling.
Some
non-recoverable alarms are generated because the instrument
detected a serious system problem that could not be resolved.
This will result in the need for the instrument to be turned OFF,
then turned ON again.
The following icons may be displayed during a non-recoverable
alarm that requires the instrument to be turned OFF:
Icon
Meaning
The procedure has been terminated. Disconnect the donor.
Turn the power OFF, then ON again to start a new
procedure.
Most of the time, an alarm screen similar to the one in the
following figure will be displayed on the touch screen.
Typical Non-Recoverable Alarm Requiring Power Cycling
The status zone displays a non-recoverable alarm code number
specific to the alarm. Touch the
mute button
to temporarily silence
the alarm’s audible tone.
Note:
The
mute button
, when touched, will silence the alarm tone
for two minutes.
Mute Button