C
HAPTER
3
–
O
PERATING CONSIDERATIONS
3-42
07-19-01-518 – January 2009
58. If desired, seal the kit on the donor line, the lines on all
solution containers, and just above the “Y” connector on the
line(s) coming from the red cell storage container(s).
59. Disconnect the storage container(s) from the kit.
60. Touch the
check button
to confirm the kit has been sealed.
Verify the
hang product prompt
is displayed.
Typical A
LYX
2RBC-LR Procedure
Information Screen with Hang Product
Prompt (Test Mode)
61. Remove the plasma, red cell and in-process containers from
their hanging scales.
62. Place one red cell storage container on the in-process scale
and the other storage container on the plasma scale.
63. Allow the weights to equilibrate. Once the weights of the
containers have equilibrated, seal each segmented tubing
next to the “Y” connector.
64. Disconnect the “Y” connector from the segment tubing,
fold each segment tubing, and place them in the respective
side slits of each storage container.
Note:
Refer to the appropriate procedure in Chapter 4, “Hang
Product to Weigh” if not performing an A
LYX
2RBC-LR
procedure.
65. Touch the
check button
to confirm that the products have
been hung for final weighing.
Hang Product Prompt
Check Button
Check Button