07-19-01-518 – January 2009
5-31
Code
Icons
Meaning
Resolution
11028
Check Pressure
Cuff
The system has detected the
pressure cuff to be inflated during a
return cycle.
- Check the cuff placement on the donor’s arm.
- Resume the procedure.
11029
Non-Recoverable
Alarm Screen
The system has detected excessive
pressure on the cassette.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
11030
Non-Recoverable
Alarm Screen
The system has detected that the
AC pump is running without the
donor pump.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact The Hotline.
11032
Non-Recoverable
Alarm Screen
The system has detected an issue
with the AC pump.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
11033
Non-Recoverable
Alarm Screen
The system has detected an issue
with the centrifuge speed.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
11034
Non-Recoverable
Alarm Screen
The system has detected an issue
with the centrifuge direction.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
11036
Non-Recoverable
Alarm Screen
The system has detected that the
AC bag is empty.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
11037
Non-Recoverable
Alarm Screen
The system has detected an issue
with the AC:WB ratio.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
11038
Non-Recoverable
Alarm Screen
The system has detected an issue
with the air detector.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
11041
Non-Recoverable
Alarm Screen
The system has detected a software
error.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
11042
&
11044
Non-Recoverable
Alarm Screen
The system has detected that too
much whole blood has been drawn
from the donor.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
11045
Remove Kit
The system has detected an issue
with the AC pump.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline
11049
Non-Recoverable
Alarm Screen
The system has detected that a path
to the red cell preservation solution
container has been opened.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
11050
Non-Recoverable
Alarm Screen
The system has detected air in the
donor line that cannot be cleared.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
11051
-
11056
Non-Recoverable
Alarm Screen
The system has detected an issue
with the power supply voltage.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.