07-19-01-518 – January 2009
4C-23
3.
Touch the
check button
to verify the clamp on the donor line
has been closed. The instrument will display the animated
kit-check icon
on the touch screen. The screen will also
provide information about the approximate time remaining
for the kit checks.
Typical Screen Indicating Kit-Checks in
Progress
WARNING: Opening the clamp between the sampling
container “Y” and the anticoagulant line “Y” after the
system self-checks and before venipuncture may cause
non-sterile air to enter the kit. If this occurs, restart the
procedure with a new kit.
Kit-Check Icon
Check Button