A
LYX COMPONENT COLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
C
HAPTER
5
–
T
ROUBLESHOOTING
07-19-01-518 – January 2009
5-27
Code
Icons
Meaning
Resolution
9001 -
9006
Non-Recoverable
Alarm Screen
The system has detected a software
error.
- If problem persists, contact the Hotline.
9009
Expired Kit
The system has detected that the
scanned expiration date on the kit is
beyond the current date in the
instrument.
- Check that the date for the kit displayed on the
instrument is correct.
- Check the expiration date on the kit and do not
use if it is expired.
- Start a new procedure using a different kit.
9012
Bar Code Error
The system has detected a problem
reading the bar code.
- Contact the Hotline.
10001 -
10007
Non-Recoverable
Alarm Screen
The system has detected a possible
issue with the interface detector.
- If problem persists, contact the Hotline.
10009
Check for Kinks
The system has detected a low flow
from the centrifuge to the cassette.
- Check the tubing from the centrifuge to the
cassette for kinks.
- Attempt to resume the procedure.
- If the problem persists, press the STOP button
to end the procedure.
-
Terminate the procedure without reinfusion
and discard the product.
10010
Non-Recoverable
Alarm Screen
The system has detected a possible
issue with the interface detector.
- If problem persists, contact the Hotline.
10011
Check for Kinks
The system has detected a possible
issue with the interface detector.
- Check the tubing from the centrifuge to the
cassette for kinks.
- Attempt to resume the procedure.
- If the problem persists, press the STOP button
to end the procedure.
-
Terminate the procedure without reinfusion
and discard the product.
10012
Remove Kit
Call for The
Hotline
The system has detected a possible
issue with the interface detector.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline.
10013
Non-Recoverable
Alarm Screen
The system has detected a possible
issue with the interface detector.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline
10014
Call for The
Hotline
The system has detected an issue
with the interface detector.
- Turn the instrument OFF, then back ON to
start a new procedure.
- If problem persists, contact the Hotline..
10100
Check Centrifuge
The system has detected that the
centrifuge door lock did not lock.
- Open and close the centrifuge door.
- Ensure centrifuge door is latched.
- Resume the procedure.