A
LYX
C
OMPONENT
C
OLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
C
HAPTER
4A
–
A
LYX
2RBC-LR
C
OLLECTION
P
ROCEDURE
4A-4
07-19-01-518 – January 2009
CAUTION:
Do not operate the instrument if the operating
display is blank or unreadable or if the instrument does not
respond to the touch screen. If the display becomes blank during a
procedure, press the STOP button to isolate the donor from the
instrument.
CAUTION:
The touch screen is the main input device for the
instrument. Do not touch the screen during operation except to
interact with the instrument.
CAUTION:
Do not switch the power OFF during a procedure,
as this will immediately end the procedure and may result in the
need for service. The instrument cannot resume from the point at
which the power was turned OFF.
CAUTION:
Failure to follow instructions prior to venipuncture
may result in dilution of whole blood sample.
CAUTION:
The battery will not resume the procedure in the case
of power loss due to accidental switching of the power switch to
OFF during a procedure.
Note:
The scales must be checked each time power to the
instrument is turned on.