O
PERATING CONSIDERATIONS
3-14
07-19-01-518 – January 2009
Scale Calibration
If a scale does not pass its initial check, a
scale fail icon
will be
displayed in the appropriate place on the touch screen.
Typical Unsuccessful Scale Check Screen
If a scale fails its initial check, continue checking the remaining
scales. Once this is complete, the touch screen will
automatically display a screen that prompts the operator to
perform a scale calibration procedure.
Typical Scale Calibration Screen
If more than one scale fails its initial check, all of the scales
requiring calibration will be displayed on the calibration screen.
The instrument will not proceed to the next step unless all
scales have passed the calibration test.
Scale Fail Icon