A
LYX
C
OMPONENT
C
OLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
C
HAPTER
7
–
S
YSTEM
S
PECIFICATIONS
7-4
07-19-01-518 – January 2009
Table 7-2 provides the operating and storage requirements for
the A
LYX
instrument.
*Mechanical and safety system operations tested and confirmed at these
temperature and humidity extremes.
**With installation of high altitude compressor assembly kit.
Table 7-2 – A
LYX
Instrument Operating/Storage
Requirements
CAUTION:
Do not turn the instrument ON if condensation is
visible on the exterior of the instrument.
CAUTION:
Do not operate the instrument outside of its
specified operating range. If the instrument has been stored
beyond its specified operating range, allow it to remain running
with the STOP test screen displayed for at least one hour to
stabilize to conditions before use.
CAUTION:
Ensure that no devices other than the pressure cuff
and bar code scanner (if used) are attached to the connection
ports on the back of the A
LYX
instrument while a donor is
connected to the A
LYX
Component Collection System.
Operating and
Storage
Requirements
Temperature Ranges
Fahrenheit
Centigrade
Operating*
60 to 90
°
F
15 to 32
°
C
Shipping and Storage
-40 to 158º F for up to
15 weeks
-40 to 70
°
C for
up to 15 weeks
Altitude Maximum
Feet
Meters
Operating Conditions
6,500
** up to 7,500
2000
** up to 2286
Humidity Ranges
Operating*
10 to 90% RH – Non-condensing
Shipping and Storage
10 to 95% RH – Non-condensing