3
–
O
PERATING CONSIDERATIONS
07-19-01-518 – January 2009
3-37
Typical Collection Volume Overlay
(Test Mode)
33. Touch the “
360” target volume button
on the overlay.
Note:
If instrument defaults have been changed or if using
another procedure, choose the lowest collection volume
available.
34. Touch the
check button
on the overlay.
Typical Donor Data Screen Complete
(Test Mode)
Note:
The instrument will not allow entry of donor or product
inputs inconsistent with configured settings of the instrument’s
Administrative Settings. An audible tone and flashing data
button indicates entry of unacceptable data input.
35. Verify the clamp to the donor line and sampling pack are
closed.
Target Volume Button
Check Button