A
LYX
C
OMPONENT
C
OLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
C
HAPTER
7
–
S
YSTEM
S
PECIFICATIONS
7-22
07-19-01-518 – January 2009
anticoagulant scale, an alarm will sound to notify the operator of
the situation.
Intravascular volume deficit (IVD) monitoring.
The
instrument’s software continuously calculates the IVD of the
donor and sounds an alarm if it is determined the IVD limit has
been reached.
Draw/return rate monitoring.
The instrument’s software
monitors draw and return rates to help maintain and stabilize
venous access to the donor and enhance donor comfort. An
alarm sounds if the donor’s flow problems continue or an
occlusion occurs in the return cycle.