PERATOR
’
S
M
ANUAL
C
HAPTER
4A
–
A
LYX
2RBC-LR
C
OLLECTION
P
ROCEDURE
07-19-01-518 – January 2009
4A-29
Typical Screen with Break Cannulae Prompt
Functionally
Closed Kit
6. Break the cannulae on all solution containers (saline (if
available), red cell preservation solution, and anticoagulant –
in any order). The cannulae are located at the inlet of each
solution container. To completely break a cannula, grasp it
in both hands, snap it at a 90-degree angle in one direction,
and then bend it in the opposite direction. To ensure proper
fluid flow, the cannula should be snapped as close to the
middle of the breakaway portion as possible.
Breaking a Cannula
CAUTION:
Ensure solution breakaway cannulae are completely
broken. Failure to completely break the cannulae may result in
restricted fluid flow.
Break Cannulae
Prompt