Once the
perform venipuncture prompt
is displayed on the touch screen,
venipuncture may be performed as follows:
1. Ensure the donor is close enough to the instrument for the
donor line to reach the venipuncture site.
2. Ensure the pressure cuff is properly positioned, inflated, and
comfortable to the donor. Refer to Chapter 3, Making
Adjustments, Pressure Cuff for information on using the cuff
pressure button.
3. Select an appropriate vein for venipuncture and prepare the
venipuncture site according to applicable standard operating
procedures.
4. Perform the venipuncture.
5. Secure the needle to the donor’s arm and touch the
check
button.
The
take samples prompt
will be displayed on the touch
screen.
Typical Take Samples Screen
WARNING: Do not open the cassette compartment door
while a donor is attached to the instrument.
6. Open the clamp on the line to the sampling pack and collect
whole blood in the sampling pack. After the desired sample
volume has been collected, close the clamp on the sampling
pack line and seal the line. After the sampling pack line is
sealed, the pack may remain with or be separated from the
kit.
Perform
Venipuncture
Perform Venipuncture
Prompt
Take Samples
Prompt
Check
Button