CAUTION:
Ensure the caps from the back of the A
LYX
instrument connection pressure ports 1 and 2 are removed
before use. These caps must be removed from the ports prior
to connecting the pressure cuff and operating the instrument.
Leaving the caps in place may result in alarms and possible
service calls.
10. If using a barcode scanner, attach the scanner cable to
port 5 on the back of the instrument.
Note:
Visually inspect the pins on the scanner and port 5 of
the instrument prior to connecting the scanner cable. Bent or
missing pins will impair the scanner’s ability to accurately scan
barcodes.
11. Gently fold out the saline scale from its stored position on
the upper-right side of the instrument.
12. Slide the leukoreduction filter fixture into its mount on the
right side of the instrument.
13. Locate the bubble level toward the front of the instrument.
Ensure the instrument is level by adjusting the
instrument’s feet until the bubble is more than half way
inside the black ring. To raise the instrument, turn the feet
clockwise; lower the instrument by turning the feet
counter-clockwise.
CAUTION:
Ensure the instrument is level prior to starting a
collection procedure.
Ensure the instrument is level each time
it is repositioned or relocated.