07-19-01-518 – January 2009
5-11
Typical Non-Recoverable Alarm Screen
An alarm screen similar to the one provided above is displayed
when the instrument has detected redness for an extended time
in the plasma line tubing coming from the centrifuge. This alarm
is usually preceded by a recoverable alarm that has already
instructed the operator to check the color of that tubing, and the
procedure continued.
WARNING: Do not attempt to reinfuse blood components
following fluid leaks or spills, when the instrument is
operating with unexpected noise or vibration, or when a
procedure is ended due to persistent redness in the plasma
tubing line coming from the centrifuge. In these instances,
the products must be discarded.
Under these circumstances, the operator should:
1. Immediately press the STOP button.
2. Isolate the donor from the instrument by closing the clamp
on the donor line. The instrument will also engage the
donor safety clamp to isolate the donor when the alarm
sounds.
3. Touch the
end button
.
4. Follow the instructions/choices on following screens that
end the procedure without fluid return. When redness in the
plasma line persists, no fluid return will be allowed.
5. Continue to follow the screen prompts to disconnect the
donor and remove the kit. Do NOT keep the product.
End Button