A
LYX
C
OMPONENT
C
OLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
C
HAPTER
4B
–
A
LYX
RBC/P
LASMA
C
OLLECTION
P
ROCEDURE
07-19-01-518 – January 2009
4B-25
When the kit checks are complete, the touch screen will
display the
break cannulae prompt
and the Kit Timer
flashing 60:00 (minutes and seconds).
Typical Screen with Break Cannulae Prompt
Note:
The operator has the option to delay starting prime.
Refer to Chapter 3, Automatic Cassette Refresh and Kit Timer
for more information.
4. Break the cannulae on all solution containers (saline (if
available), red cell preservation solution, and anticoagulant –
in any order). The cannulae are located at the inlet of each
solution container. To completely break a cannula, grasp it
in both hands, snap it at a 90-degree angle in one direction,
and then bend it in the opposite direction. To ensure proper
fluid flow, the cannula should be snapped as close to the
middle of the breakaway portion as possible.
Breaking a Cannula
Break Cannulae
Prompt