CAUTION:
Ensure solution breakaway cannulae are
completely broken. Failure to completely break the cannulae may
result in restricted fluid flow.
5. Touch the
check button
to verify all cannulae have been
broken. The animated
prime icon
will be displayed on the
touch screen.
WARNING: Remove and discard the kit if any fluid leaks
are observed after it is primed. If you detect a leak in the kit
at any time, stop the procedure, do not reinfuse, and discard
the blood product.
Note:
Donor information may be entered any time after the
operator ID and kit information is entered. This may occur
either before or after the disposable kit has been installed on the
instrument or during the prime sequence as described in the
following section.