A
LYX
C
OMPONENT
C
OLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
C
HAPTER
4D
-
2RBC
S
INGLE
U
NIT
R
ECOVERY
P
ROCEDURE
07-19-01-518 – January 2009
4D-7
If the ability to both Return fluid to the donor and Process the
product are options, the confirmation overlay will ask the
operator to confirm the choice to return fluids to donor and
process the product. After touching the
Yes button
, the instrument
will then proceed to reinfuse any fluids collected that exceed the
minimum volume needed to allow for processing a product.
After the
Yes button
is touched, the
disconnect donor prompt
is
displayed and the instrument will automatically proceed to collect
the red cells from the in-process container
.
Typical Disconnect Donor – Processing IP Container –
2RBC-LR (SUR)
To disconnect the donor, follow these steps:
1. Close the clamp on the donor line.
2. Remove the pressure cuff from the donor’s arm.
3. Remove the needle while simultaneously sliding it into the
needle protector. Ensure the needle is completely covered
by the needle protector.
4. Apply pressure to the venipuncture site to stop blood flow.
5. Follow institutional guidelines for venipuncture site care and
release the donor according to applicable standard operating
procedures.
Disconnecting
the Donor
Disconnect
Donor Prompt
Содержание 4R5707
Страница 2: ......