A
LYX
C
OMPONENT
C
OLLECTION
S
YSTEM
O
PERATOR
’
S
M
ANUAL
C
HAPTER
4C
–
A
LYX
2RBC-NLR
C
OLLECTION
P
ROCEDURE
4C-24
07-19-01-518 – January 2009
When the kit checks are complete, the touch screen will display
the
break cannulae prompt
and the Kit Timer flashing 60:00
(minutes and seconds).
Typical Screen with Break Cannulae
Prompt
Note:
The operator has the option to delay starting Prime.
Refer to Chapter 3, Automatic Cassette Refresh and Kit Timer
for more information
.
4. Break the available cannulae on all solution containers
(saline, red cell preservation solution, and anticoagulant – in
any order). The cannulae are located at the inlet of each
solution container. To completely break a cannula, grasp it
in both hands, snap it at a 90-degree angle in one direction,
and then bend it in the opposite direction. To ensure proper
fluid flow, the cannula should be snapped as close to the
middle of the breakaway portion as possible.
Breaking a Cannula
CAUTION:
Ensure solution breakaway cannulae are
completely broken. Failure to completely break the cannulae may
result in restricted fluid flow.
Break Cannulae
Prompt