Power Transformer
In case of repair, replace the complete power transformer.
14.5.1
Testing the Power Transformer Output Voltage
• Switch control box off using ON/OFF switch (ON/OFF switch is located on rear
panel of control box).
• Unplug power cord from AC outlet.
• Check mains fuses of the control box (mains fuses are located on rear panel of
control box).
• If mains fuses are OK, proceed as follows:
• Remove control box from trolley (see control box chapter "
• Remove lateral screws
1
from control box
2
and lift top cover
3
in direction of
A
and away from control box
2
.
• Place top cover
3
of control box
2
aside.
Fig. 13:
Removing top cover of the control box
Electrostatic discharge can damage electrostatic sensitive
devices. Use a static-dissipative mat and a wrist strap when
handling electrostatic sensitive devices.
1..2..
Menu
PEEP
IPPV
man./
spont.
Pmax
TI:TE
TIP:TI
f
IPPV
V
T
IPPV
Paw [mbar]
FiO2
MV
Paw
60
40
20
0
24
7.9
28
0
80
21
12.0
3.0
40
8
1
1
A
2
3
Содержание Fabius 4116029
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...