Fabius
®
Repair Instructions
3-112
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_
repa
ir
instr
uc_c.
fm
19.2
Testing the S-ORC
19.2.1
Mechanical Pre-Adjustment
• Adjust adjusting screw
1
for O
2
such that it protrudes about 1 mm from valve
element
2
.
Fig.100:
S-ORC, adjusting the O
2
flow
• Adjust adjusting screw
3
for N
2
O such that it protrudes about 1 mm from valve
element
4
.
Fig.101:
S-ORC, adjusting the N
2
O flow
• Adjust distance between upper edge of valve elements
4
and
5
(see Figure
above) and housing to approximately 10 mm.
N
2
0
0
2
A
E
E
A
1
2
N
2
0
0
2
A
E
E
A
3
4
5