4-29
24.3
Flow sensor zeroing:
(With SW > 2.5 check that gas type is set
to AIR)
Move to STANDBY. Detach hoses of flow
sensor from control box.
Call-up zeroing page by pressing
p
Zero flow sensor.
Zeroing successful..
Move to IPPV and set a fresh-gas flow of
3 L/min. Reconnect hoses, connect flow
sensor to ventilator output as per drawing.
Set
pmax
= 70mbar
V
T
=500
mL
f
IPPV
= 6/min
TI:TE
= 1 : 1
TIP:TI
= 10 %
PEEP
= 0mbar
Measured flow = 420 - 580 mL
Install flow sensor.
Upgrade unit ready for operation.
24.4
IPPV ventilation
Setting:
APL valve set to IPPV
Make the following control box settings:
Pmax = 70 mbar
Vt = 380 mL
fippv = 12 1/min.
TI:TE= 1:1
TIP:TI = 10%
PEEP = 0 mbar
Test thorax connected
Press IPPV button
Display: IPPV mode?
Press
button
O
Breathing hose0,3 - 1 m
ISO socket M 25647
Flow sensor
V
°
Содержание Fabius 4116029
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...