Fabius
®
Fault-Cause-Remedy
6-3
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec6_f
aul
tca
usrem
.fm
FLOW INOP
The flow sensor has not
been calibrated.
The flow sensor is faulty.
The flow sensor has been
replaced but not
calibrated.
The flow sensor hose is
faulty.
Calibrate the flow sensor.
Replace the flow sensor and
calibrate it.
Calibrate the flow sensor.
Replace the flow sensor hose and
calibrate the flow sensor.
MEMORY ERROR
Internal data loss.
Switch Fabius off and on.
Inform DrägerService, if
necessary.
MV HIGH
The upper alarm limit for
the minute volume has
been exceeded.
The flow sensor has not
been calibrated.
The flow sensor is faulty.
Calibrate the flow sensor (see
Instructions for Use).
Calibrate the flow sensor.
Replace with new flow sensor and
calibrate new flow sensor.
MV LOW
The minute volume has
decreased below the
lower alarm limit.
The hose is
blocked/kinked.
There is a leak in the
breathing system.
Reduced volume during
pressure limitation.
Reduced lung compliance.
The flow sensor has not
been calibrated.
The flow sensor is faulty.
Check the minute volume.
Check the hose system.
Check the breathing system.
Leak test the breathing system.
Adjust the ventilation parameters
correctly.
Calibrate the flow sensor.
Replace with new flow sensor and
calibrate new flow sensor.
Fault
Cause
Remedy
Содержание Fabius 4116029
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...