Fabius
®
Repair Instructions
3-116
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_
repa
ir
instr
uc_c.
fm
19.2.7
Fine Adjusting O
2
Concentration (O
2
/N
2
O ratio)
• Fully open N
2
O flow control valve on flowmeter block.
• Adjust O
2
flow control valve to 1 L/min.
N
2
O flow should be 3 L/min (±0.3 L/min). O
2
concentration at fresh-gas outlet should
be 25 vol.% O
2
(±3 vol.% O
2
).
• If N
2
O flow is not 3 L/min (±0.3 L/min) or if O
2
concentration at fresh-gas outlet is
not 25 vol.% O
2
(±3 vol.% O
2
), proceed as follows:
• Use adjusting screw
1
on N
2
O side to adjust above mentioned values.
Fig.104:
S-ORC, adjusting the N
2
O flow
i
Make sure O
2
sensor is functioning properly and has been
calibrated. Check O
2
concentration using an O
2
analyzer (accuracy:
1 vol.% O
2
).
N
2
0
0
2
A
E
E
A
1
Содержание Fabius 4116029
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...