4-30
Unit starts ventilation.
Drive runs quietly and smoothly
24.5
Display V
te
= 300 - 450 ml
24.6
Peep pressure
Steady-state control after max. 10
breaths.
Adapt pressure warning limits if
necessary
Press PEEP parameter button.
Display indicates ventilator settings.
Set the following Peep values.
Setting Test
value:
0
Ppeep = 0
+
2 hPa
10
Ppeep = 10
±
2 hPa
15
Ppeep = 15
±
3 hPa
Set Ppeep = 0 hPa.
24.7
Pmax check:
Seal off Y-Piece. Connect pressure gauge
to P
AW
port.
Set following Pmax values:
Setting Test
value:
25
P = 25
+
5 hPa
40
P = 40
±
5 hPa
60
P = 60
±
5 hPa
70
P = 70
±
5 hPa
24.8
APL valve check:
Set Fabius to „MAN“ .
Increase fresh-gas flow to 5 L/min.
Y-piece is sealed.
Set APL valve to „MAN“, 30 mbar.
Pressur display:
27 -33 hpa
Set APL valve to „IPPV/SPONT“
Pressure display:
0-3mbar
Set flow to zero, unit set to STANDBY.
O
O
O
V
V
V
V
V
V
V
V
V
Содержание Fabius 4116029
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...