Fabius
®
Repair Instructions
3-12
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_
repa
ir
instr
uc_c.
fm
7.3.2
Flow Curve (software 2.00), Page 2 (Keycode “2”)
Flow curve
(Quit key: toggles between 1:1
and 1:6 resolution)
Measured tidal volume in mL
(normal operating range: 0 to 9999 mL)
Vte XXXX mL
Expiratory VT during inspiration (in mL, ideally: 0)
(normal operating range: 0 to 9999 mL)
Vtix XXXX
mL
Expiratory peak flow in L/min
(normal operating range: 0 to 150 L/min)
Flep XXX L/min
Tidal volume corresponding to ventilator’s I/E
phase (in mL)
(normal operating range: 0 to 9999 mL)
Vtev XXXX
mL
Plateau pressure in hPa
(normal operating range 0 to 99 hPa)
Plat XX
hPa