Fabius
®
Fault-Cause-Remedy
6-2
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec6_
fau
ltcausr
em
.fm
CHECK VENTILATOR
The ventilator is not
assembled correctly.
Check the diaphragm and close
the housing of the patient system.
Select standby mode and switch
back to IPPV mode.
CONT HIGH PRESS
The APL valve is in the
position MAN during IPPV.
Switch APL valve to
IPPV/SPONT.
DEVICE TEMP HIGH
The device temperature is
too high.
Check for external heat sources.
Check the fan.
EX-VALVE LEAKAGE
There is a leak in the
expiratory valve.
Check the expiratory valve.
FiO
2
HIGH
The O
2
flush is in use, the
fraction of inspired O
2
exceeds the upper alarm
limit.
Check the O
2
setting on the
flowmeter block.
Check the concentration setting
on the anesthetic vaporizer.
FiO
2
INOP
The O
2
sensor has not
been calibrated correctly.
The O
2
sensor has been
replaced but not
calibrated.
Spent O
2
sensor.
Faulty O
2
sensor cable.
Calibrate the O
2
sensor.
Calibrate the O
2
sensor.
Replace the O
2
sensor and
calibrate it.
Replace the O
2
sensor cable and
calibrate the O
2
sensor.
FiO
2
LIMIT < 21%
The lower limit FiO
2
is
between 18% and 20%.
Set appropriate FiO
2
lower limit.
FiO
2
LOW
The FiO
2
is below the
lower alarm limit.
Check the O
2
supply.
Check settings on flowmeter
block.
Fault
Cause
Remedy
Содержание Fabius 4116029
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 8: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 68: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 234: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 318: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 324: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 326: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...