Fabius
®
Repair Instructions
3-32
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_
repa
ir
instr
uc_c.
fm
• Plug power cord into AC outlet.
• Switch control box on.
• Using a multimeter, measure the power transformer output voltage (it should be
approx. 40 VAC).
• If the power transformer output voltage is not approx. 40 VAC, check fuses, if
fuses are OK, replace power transformer (see "
Dismounting/Replacing the Power
Fig. 14:
Power Supply PCB, measuring the power transformer output voltage
X20
X 4
X 2
T4L
T2,5L
X8
X6
T1,6L
T3,15L