Safety Valve of the Patient System
If the pressure limit (Pmax) is exceeded considerably, or if the pressure limit control
fails, the patient system safety valve limits the gas pressure. This safety valve is
permanently set to a working pressure of 60 hPa (mbar) to 80 hPa (mbar).
Fig. 31:
Sectional view of the safety valve
valve disc
washer
screw
spring