7.1.1
Function Principle of the Ventilator
The ventilator consists of a piston and a cylinder. The recirculating ball screw driven
by the motor moves the piston. When the piston moves back it slides completely over
the motor. The incremental encoder determines the number of motor rotations and
transmits the corresponding signal to the microprocessor. The rolling seal is attached
to the piston. The bag-type rolling seal encloses the inspiratory volume.
Fig. 26:
Sectional view of the ventilator
patient system
rolling seal
recirculating ball screw
motor
piston
cylinder
bag-type
rolling seal