Fabius
®
Repair Instructions
3-104
Cop
yr
ight
2000 Dr
aeger
M
edical,
In
c.
sec3_
repa
ir
instr
uc_c.
fm
18.1.2
Dismounting/Replacing the Flowmeter Block
• Switch control box off using ON/OFF switch (ON/OFF switch is located on rear
panel of control box).
• Unplug power plug from AC outlet.
• Disconnect device from gas supplies.
• Remove fixing screws
1
from flowmeter block
2
.
• Pull flowmeter block slightly out making sure that shield
3
of flowmeter block
2
does not fall down.
• Slide shield
3
out of the flowmeter block
2
.
• Place shield
3
shield aside.
Fig. 88:
Removing the flowmeter block
• Pull flowmeter block slightly out.
Fabius
S-ORC
Oxygen-Ratio-Controller
1
2
1
3